3) PROGRAM (SOFTWARE) OPTION
(1) Press mode key, DP+CAL LED turn on at the same time.
mode
Adjustment of peak
hold mode
mode
Parity of
Display
Tracking
mode of
zero point
Average
value
display of
Signal
7. ALARM OUTPUT (RELAY OUTPUT)
High and low relay are a and b contact respectively.
Contact capacity is maximum AC 250V/5A and DC 30V/5A.
Relay contact operating time is maximum (1/Sampling Time) mm Sec
* Output of transistor NPN open collector is option.
mode
HIGH
high data > setting value
RELAY
high data ≤ setting value
LOW
low data ≤ setting value
RELAY
low data > setting value
display
peak hold on : display maximum measurement data
0 x x x
peak hold off : display present measurement data..
reset : stored maximum measurement data is removed.
peak hold on(only) : display and reform
maximum measurement data
1 x x x
peak hold off : stored maximummeasurement data is
removed automatically.
display
standard type.
x 0 x x
ex) +20mA : display 1000.0
display (+) to (-)
x 1 x x
signal and display indicate opposite.
ex) +20mA : display -760.0
standard type.
x x 0 x
Display 0 : trace minute signal to 0
Display 0 : display minute signal.
x x 1 x
Use in case of leakage measurement.
x x x 0
display measurement signal immediately.
display average value of measurement
x x x 1
signal (7 times sampling).
setting
description
description
relay
ON
OFF
ON
OFF