Astrea Evolve Panduan Pengguna - Halaman 14
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan Industri Astrea Evolve. Astrea Evolve 19 halaman. 350mm column instructions for replacing the wetted flowpath using the column refresh kit
Step 2-Reassembly and Refresh of the fixed assembly.
The fixed end flowcell is supplied complete with the process port in place.
1.
Place the bottom flowcell on the backing plate using the alignment indicators to ensure flow cell is in the correct
orientation.
2.
Secure the flow cell to the column base plate using the bottom flowcell bolts.
3.
Locate the bed support seal into the inner most groove on the flowcell.
4.
Locate the fixed flange seal into the outer most groove on the flowcell.
5.
Locate the bed support with the smooth side uppermost and secure using the bed support retaining bolt. Do not
overtighten this bolt.
Note: Always use clean dry gloves when handling bed supports.
6.
Connect the pipespool elbow to the flowcell and secure in place by refitting the pipespool clamp top.
Figure 14 Reassembly of Column Bottom Flowcell and Fixed end
®
Evolve
350mm Column: Instructions for Replacing the Wetted Flowpath using the Column Refresh Kit
14