Cognionics Quick-30 Pembersihan dan Perawatan - Halaman 2

Jelajahi secara online atau unduh pdf Pembersihan dan Perawatan untuk Peralatan Laboratorium Cognionics Quick-30. Cognionics Quick-30 4 halaman.

Cognionics Quick-30 Pembersihan dan Perawatan
 
 
 
Usage:  
The Quick-20 is designed to be a robust, durable headset suitable for recordings in real-world
 
environments. To keep your Quick-20 functioning optimally, take note of the following items: 
 
1. Avoid excessively bending the legs or earband - The Quick-20 relies on a flexible
material of the legs to deflect conform across the variations in human head size and
shapes. Do not grab and deliberately bend the legs or band beyond the range that
would normally be encountered in usage. 
2. Do not pull - The Quick-20 relies on the flexible material of the legs to deflect conform
across the variations in human head size and shapes. Do not grab and deliberately
bend the legs beyond the range that would normally be encountered in usage. 
3. Never connect the Quick-20 to a standard USB port - The microUSB port on the
Quick-20 is designed to work only with Cognionics accessories, which provide
electrical isolation to protect the headset and subject. Connecting the Quick-20 to a
PC (without the wired data dongle) or USB wall charger may damage the headset. 
 
Cleaning: 
 
The headset should be cleaned with the provided electric brush (Fig. 1).
 
Apply just enough Clear Purell or 70% isopropyl alcohol to dampen the
 
bristles of the electric brush and begin cleaning the headset (Fig. a) 
 
Applying too much isopropyl alcohol may cause the headset to
malfunction until it is completely dried. Using any liquid other than Clear
Purell or isopropyl alcohol may leave residue and cause the headset to
 
malfunction and give incorrect impedance values. 
 
Begin cleaning the headset with the sensors still inserted. The headset is designed to be
 
cleaned this way. Figure b shows proper technique for cleaning the sensors with the electric
 
brush.  
 
 
   
       
 
 
   
 
   
 
 
 
 
 
 
   
 
 
 
   
 
 
 
   
 
 
   
 
 
 
 
 
 
 
 
 
 
 
     
 
 
 
 
 
 
 
 
   
NOT APPROVED FOR MEDICAL USE