5.13 Barometric reading (ASC-400-1 only):
1. Connect a Barometric reference source to ASC-400 barometric sensor using a
tube (5 mm inner diameter / 8 mm outer diameter). ASC-400 barometric sensor
is located below the batteries underneath a "Calibration void" label.
2. Type "13" in the Calibration menu.
Prompt = "Apply 0.95 bar absolute - Press <enter> when stable"
3. Apply 0.95 bar absolute to the ASC-400 and press "enter" when stable.
Prompt = "Apply 1.05 bar absolute - Press <enter> when stable"
4. Apply 1.05 bar absolute to the ASC-400 and press "enter" when stable. ASC-
400 now returns to the Calibration menu.
5. Apply a new "Calibration void" label over the ASC-400 barometric sensor after
calibration.
5.14 System Calibration Date:
1. Type "14" in the Calibration menu.
2. Follow the menu prompts.
Prompt = "Calibration Date: 2014 9 5 13 1 24"
Prompt = "Enter new date <year> <month> <day> <hour> <minute> <second>:"
Example: 2014 09 05 13 55 14
5.15 Barometric Pressure Calibration Date:
1. Type "15" in the Calibration menu.
2. Follow the menu prompts.
Prompt = "Calibration Date: 2014 9 5 13 1 24"
Prompt = "Enter new date <year> <month> <day> <hour> <minute> <second>:"
Example: 2014 09 05 13 55 14