APPENDIX A
After Service Test Procedure
The following test should be performed after any service to ensure proper operation of the system.
Check if the repaired module meets the specifications as described in the Service- or Instruction
Manual.
Carry out the following Electric Safety check:
- Earth Leakage Current general.
- Enclosure Leakage Current
- Patient Leakage Current
Above tests should be performed in accordance with the requirements of IEC601-1 clause 19.
30200300E
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
17/17