BOQU PHS-1705 Panduan Pengguna - Halaman 9
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Alat Ukur BOQU PHS-1705. BOQU PHS-1705 20 halaman. Laboratory ph meter
measurement end point A is the default setting for the instrument. When the
electrode output is stable, the display is automatically fixed and the pH of the
sample solution is displayed.
Press and hold the "OK" button, can switch between automatic and manual
measurement mode end. To manually measure an end point, press the "OK"
button to display the value.
pH and mV measurement stability based on: If the signal changes within 6
seconds is not greater than 0.1mV, the instrument will reach the end of the
measurement.
To view the mV value during the pH measurement, simply press the "Mode"
button to execute the mV (ORP) measurement, according to the same steps as the
pH measurement.
2.3 Temperature measurement
To improve accuracy, we recommend using a temperature probe or an electrode
with a built-in temperature probe. When using the temperature probe, the screen
displays the ATC symbol and the sample temperature.
Note: This instrument uses only the NTC30k temperature probe.
2.4 Parameter setting
Press the "Setup" key to enter the parameter setting mode, press "Setup"/"Mode"
to select the desired setting parameters,
Press "Exit" to exit the parameter setting mode, return to the measurement mode;
Press "OK" to enter the corresponding parameter setting page, press the
- 6 -