ALARMS & INDICATORS
OCC/EMP
DOSE COMP →Dose limit has been reached.
(4 beeps, flashing indicator, 2½ minute repeats)
FLO FAST
(Loud 3 tone beep)
LOW BAT
To Reset Alarms
1
2
3
→Pathway occluded
→Container empty
→Drip chamber improperly installed, or
→Pump inactive for over 2½ minutes
→Tubing not on rotor, or
Drop sensor blocked over 6 seconds.
→Low battery. Plug into AC outlet
1. Press RUN/STOP
2. Correct alarm condition
3. Press RUN/STOP to resume feeding