Daily:
Check the filter glass for damage and dirt; check the housing for damage.
Every 400 operation hours / each 2-3 years in private operation:
Replace the lamps.
Every 2 years: Maintenance by our Customer Service or a technician authorized by us
(applies only to commercial users but is recommended to comply with the
Medical Devices Act).
5.1.
C
LEANING THE
Disconnect the unit by unplugging the power-pack!
Never wipe the hand lamp or the filter glass with a dry cloth. To clean dusty or dirty surfaces,
wipe with a damp soft cloth or sponge (add a little dishwashing liquid to warm water).
Anti-static plastic cleaners are also suitable.
Never clean with concentrated disinfectant or solvent such as Sagrotan, Lysoform,
methyl alcohol or any other liquid containing alcohol.
5.2.
C
LEANING AND
The comb attachment (Art.-No. 995200) is skin-friendly and hygienic. Its poreless surface
makes it difficult for dirt and bacteria to accumulate.
To clean and disinfect the comb, see Section 5.2.
At 25 ° C the comb is resistant to the following chemicals:
•
-
Baktolan, up to 5 %
-
Chinosol, up to 1 %
-
Elmocid Gamma, up to 2 %
-
Lugol's solution
-
Mefarol, up to 1 %
-
Merckojod, up to 1 %
I
NFORMATION
H
L
AND
AMP
D
ISINFECTING THE
-
-
-
-
-
-
11
dermalight
B
EDIENUNGSANLEITUNG
C
A
OMB
TTACHMENT
Merfen
Perhydrol
Perodin
Sagrotan, up to 2 %
Hydrogen peroxide, up to 40 %
Zephirol, up to 5%
®
80 uva
O
I
PERATING
NSTRUCTIONS