Dr. Mach 170 120 1300 Petunjuk Pemasangan - Halaman 12
Jelajahi secara online atau unduh pdf Petunjuk Pemasangan untuk Peralatan Medis Dr. Mach 170 120 1300. Dr. Mach 170 120 1300 19 halaman. Stand lamp
Stand for Mach M2
5.3 Maintenance of the securing segment
•
Dismount the lamp head in reverse order of mounting (see chapter 3.2.2).
•
Check the thickness of the segment. It must be at least 1,5mm. In case the thickness of the
segment is less than 1,5mm it has to be changed.
•
Grease the segment and the coupling journal of the lamp.
•
Mount the lamp to the spring arm, as described in chapter 3.2.2.
Electrical Data
Rated voltage
depending on model
Maximum power
depending on model
Nominal frequency
Fuses
5 x 20 mm, time-lag T
Rated voltage prim.
120 V
230 V
100 V to 127 V
213 V to 247 V
120 V
230V
The device is not suitable for use in combustible blends of anaesthesia agents with air or for
use in combustible blends of anaesthesia agents with oxygen or nitrous oxide.
The device is designed for continuous operation.
59860007
6. Data
6.1 Technical data
100 V / 120 V / 230 V
60 W / 90 W
50 / 60 Hz
Rated power
of the transformer
60 VA
60 VA
60 VA
60 VA
90 VA
90 VA
Edition 04
Dr. Mach
Lamps and Engineering
Fuse rating
1,0 A
0,5 A
1,25 A
0,63 A
1,6 A
0,8 A
22.07.2009 / Bak Page 12/19