Dr. Mach LED 115 Petunjuk Penggunaan Manual - Halaman 9
Jelajahi secara online atau unduh pdf Petunjuk Penggunaan Manual untuk Peralatan Medis Dr. Mach LED 115. Dr. Mach LED 115 15 halaman. Spot light with led technology
Mach LED 115
The light Mach LED 115 is equipped with a clamping lever on the power supply housing. After
the mounting of the light with the mobile stand or the other fixation designs this clamping lever
must be adjusted according to the customer's requirements.
Preventive maintenance of the light must be done every two years. This includes a technical
and mechanical check-up.
Remark: Before doing any maintenance work on the light the light must be turned off and
disconnected from mains. Please protect the light against switching on.
5.1 Periodical maintenance work
The following maintenance work / tests has / have to be done every six months:
check on defects in paint work;
check on fissures at plastic parts;
check on deformation of the suspension.
The following maintenance work / tests has / have to be done once a year:
check the function;
electrical safety test;
check the suspension.
59060001
5. Maintenance
5.2 Settings at the light arm
Adjusting the clamping lever
In case the movement of the light head is tight or
it doesn't hold its position anymore, the braking
efficiency of the clamping lever on the power
supply housing must be adjusted.
5.3 Settings at the mobile stand
One-handed height adjustment
Adjusting the spring force at these stands is
not possible.
It is usually not necessary to adjust the spring
force at stands with one handed height ad-
justment.
Edition 01
Dr. Mach
Lamps and Engineering
02.08.2012 / Bak Page 9/15