Drawell B104 LED Panduan Pengguna - Halaman 15

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan Laboratorium Drawell B104 LED. Drawell B104 LED 19 halaman. Biological microscope

Drawell B104 LED Panduan Pengguna
hours continuously, better cut off the microscope about 30 minutes.
5.4 Replacement of fuse (Shown in Fig.8)
5.4.1 Cut off power of microscope, and pull out the plug.
5.4.2 Unscrew fuse cap in the back of base, take out old fuse.
5.4.3 Replace a new fuse, then screw the fuse cap.
5.5 Stop to use microscope, please cut off power, cover the dust cover, and place
it in a cool and dry environment.
5.6 B-LED biological microscope
5.6.1LED illumination specification:
Available LED using time: Above
10,000 hours
Working Voltage: 3.1V
Maximal working electrical current:
30mA
Maximal power dissipation: 120mW
Luminous intensity: Above 6cd
5.6.2LED bulb replacement:
Cut off power: Pull out the power plug off socket, and confirm power cut off
Put out bulb cover: Put out the bulb cover off the up-cover of the baseLED
bulb replacement: Pull up the LED bulb with plug and out off the bulb socket,
then insert the replacing bulb with its plug into the bulb socket
13
Bulb Cover
LED bulb
Bulb plug
Bulb socket