Drawell MDF-40V50 Panduan Pengguna - Halaman 5
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan Medis Drawell MDF-40V50. Drawell MDF-40V50 17 halaman. Medical freezer
I. Main performance and characteristics of the
product
Thermal insulating door and secondary seal door are adopted for use convenience. The loss of
refrigeration capacity is reduced and the energy consumption is saved.
Air-cooled condenser is adopted which is provided with good heat dissipation and fast refrigerating speed.
Stainless steel inner and tube-in-sheet evaporator are adopted. Things can be put in or taken out easily.
The refrigeration effect is good and the freezer can be cleaned easily.
Import microcomputer temperature controller is adopted for accurate temperature control. The
temperature inside the freezer is displayed digitally. With elegant appearance, the freezer can be used
conveniently. Users can regulate and check the inside temperature and the working state without opening
the door.
The microcomputer control is more accurate and safer.
The system is equipped with a safety protection device for ensuring the safe and reliable operation of the
system.
The refrigerating system all adopts imported parts which provide good reliability and long service life to the
system.
II. Precautions
The sign of "
products, or alarm users should pay special attention to this.
:It is strictly prohibited to unpack the outer packing case for transportation.
The ultra low temperature freezer should be used under the environmental temperature of 10℃~32℃. It is
suggested to use it under the environment with air conditioning.
:The slant angle of the body of the ultra low temperature freezer should not be larger than 45° in
transportation to avoid faults of the compressor or the system.
:After unpacked, the freezer should be placed on flat and firm ground near to the power socket.
:Do not put it near the heat source and avoid the sunshine.
: The ultra low temperature freezer should be stored in a dry and drafty room without corrosive gas
surrounding.
: If the ultra low temperature freezer has to be stopped for a long period, it should be cleaned inside and
outside and then covered properly with a plastic housing.
:The power supply is 220V/50Hz. If the voltage is instable, please equip a power regulator which is
above 3000VA and please use it alone.
: It is necessary to provide individual one-phase three-hole socket and proper fuse for the ultra low
temperature freezer.
: The ultra low temperature freezer must be provided with reliable independent ground wire. The ground
" means the place is dangerous to human bodies, or the use pattern is harmful to
3