BD Alaris 800TIG2GBD1 Panduan Referensi Cepat - Halaman 9
Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Pompa Air BD Alaris 800TIG2GBD1. BD Alaris 800TIG2GBD1 12 halaman. Volumetric pump
An alarm indicates that the infusion has stopped.
Check the display for an alarm symbol.
Press
to silence the sound. (Exceptions are
c
silenced)
When the cause has been rectified, press
All alarms will stop the infusion except for
where the infusion status will remain the same state as prior to
the alarm.
Alarm / Condition
J
Air-in-line
Upstream occlusion
K
Internal battery depleted.
L
Door open
M
System fault
N
Flow error
Flow sensor attached to secondary
infusion set.
O
Flow sensor connection error.
P
Downstream occlusion
Q
Incorrect infusion set, infusion set
incorrectly loaded or infusion set worn.
and
M
K
key to resume infusion.
d
F
Action
• Disconnect the infusion set from the
IV access device and use the prime
function.
• Remove occlusion/air and restart the
infusion.
• Connect pump to AC power.
• Ensure pump is switched on to charge
internal battery.
• Close door and restart infusion.
• Switch pump off.
• Remove pump from service and have
the pump inspected by a qualified
service engineer.
• Close in-line clamp.
• Ensure tubing is loaded correctly.
• Ensure ample fluid is in container.
• Clear blockage. Close pump door.
• Restart infusion.
• Ensure flow sensor is attached to
primary infusion set.
• Connect flow sensor to pump and
restart infusion. Ensure fluid in drip
chamber is not above fill line.
• Remove pressure in the infusion set to
prevent a post occlusion bolus to the
patient.
• Remove the cause of the blockage.
• Restart the infusion.
• Use infusion set as specified in DFU.
Ensure infusion set is loaded correctly
into pump.