AND TM2441-EX Buku Petunjuk - Halaman 3
Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Monitor Tekanan Darah AND TM2441-EX. AND TM2441-EX 3 halaman. Ambulatory blood pressure monitor
Instruction Manual Page 53 8.3.1. A-BPM Items and Parameters
The following changes have been made.
#3 : Example for automated measurement.
If the start time is specified and operation time is set
to "OFF", when the AUTO switch is set "ON", A-
BPM starts at the preset start time and continues until
the AUTO switch is set to "OFF". If the AUTO
switch is set to "ON" again, A-BPM starts at the
preset start time.
If the operation time is specified and the AUTO switch
is operated during A-BPM, A-BPM is continued for
operation time form this operation.
#4
: Example for automated measurement.
If the start time is set to "OFF" and operation time is
specified, when the AUTO switch is set to "ON", A-
BPM is started for the operation time. If the AUTO
switch is set to "OFF" during A-BPM, A-BPM stops. If
the AUTO switch is set to "ON" again, A-BPM is
performed for the operation time.
If the start time is specified and the AUTO switch is
operated during A-BPM, A-BPM is stopped and is started
at the preset start time next day.
Instruction Manual Page 56 8.3.2. A-BPM Program Examples
The following notes were added to
NB. Automatic blood pressure measurement will continue even if turned "OFF" and "ON" again during
automatic measurement.
The following notes were added to
NB. Automatic blood pressure measurement will not be performed once the operation time has been
exceeded even if turned "OFF" and "ON" again during automatic measurement.
Wrong
Example 2 Automatic measurement
"
Example 3 Automatic measurement
"
#3: If the start time for automatic measurement is specified
and the operation time is set to "OFF", turning "ON"
the AUTO switch will start automatic blood pressure
measurement from the start time set and measurement
will continue until the AUTO switch is turned "OFF".
If the AUTO switch is turned "ON" again, automatic
measurement will continue.
If the operation time is set, automatic blood pressure
measurement will be performed from when the
switch is initially turned "ON" until the operation time has
passed even if the
through measurement.
#4: If the start time for automatic measurement is set to
"OFF" and the operation time is set, turning "ON" the
AUTO switch will implement automatic blood pressure
measurement after initial automatic measurement until
the set operation time has passed. If the
switch is turned "OFF" midway through automatic blood
pressure measurement, automatic measurement will
stop. If the
automatic measurement will be performed until the set
operation time has passed.
If the start time is set, operating the
midway through automatic blood pressure measurement
will stop automatic measurement and turning it "ON" again
will resume measurement.
Addition
Addition
3
Correct
AUTO
switch is operated midway
AUTO
switch is turned "ON" again,
"
"
1WMPD4004496
AUTO
AUTO
AUTO
switch