Advanced Instruments 3D3700 Penggantian
Jelajahi secara online atau unduh pdf Penggantian untuk Peralatan Laboratorium Advanced Instruments 3D3700. Advanced Instruments 3D3700 2 halaman. Sample probe
Reference: Use this instruction with replace-
ment part 3D3700 (stainless
steel) or 4D3102 (non-metallic),
and any of the following models
of Advanced Instruments, Inc. or
Fiske Associates
cryoscope:
Advanced Instruments Inc.
4250 Cryoscopes (Serial Suffix E - W)
4D3 Cryoscopes (Serial Suffix E - W)
3250 Osmometers (Serial Suffix C - T)
3D3 Osmometers (Serial Suffix C - T)
Fiske Associates
Mark 2 Cryoscopes
Mark 3 Osmometers
Tools Needed: Phillips screwdriver, flat-
bladed screwdriver, probe
alignment tool and instruc-
tions (included).
CAUTION: Improper connections may cause
damage to the instrument.
Instruction:
1. Place an empty sample tube in the freez-
ing chamber to catch any extraneous
material that might fall in.
Sample Probe Replacement
3D3700
®
osmometer or
A
For additional information or technical assistance,
please contact Advanced Instruments Hot-Line
Service Center (U.S. 1-800-225-4034, outside North
America +US 1-781-320-9000).
2. Turn off the power and unplug the instru-
ment.
3. Remove the two screws from the head
cover (A) and lift off the head cover.
4. Unplug the sample probe connector from
the head transition board (B).
5. Loosen the two mandrel setscrews (C).
6. Lower the probe and mandrel together,
down over the stir/freeze wire.
7. Push the connector and lead wires of the
new sample probe up through the hole in
the sample head, and raise the mandrel so
that the stir/freeze wire extends down
through the stir/freeze wire channel in the
mandrel.
®
(3255/4D35/MK05 Service Manual)
B
C
3DP700 Rev2
Page 1 of 2