OPERATING INSTRUCTIONS
Refer to "Conditions for Testing" on the page 6 before conducting a Screening or Standard Test.
SCREENING TEST
1. Attach a sampling cup onto the sample port and select the Screening Test button.
2. When the screen displays "Please Blow", take a deep breath and blow into the sampling cup until the BEEP
tone stops.
3. An "Alcohol Detected" or "No Alcohol" result will display on the screen.
4. If the screen shows "Alcohol Detected", use a mouthpiece to perform a Standard Test to obtain a specific
alcohol reading.
5. The operator can get an immediate Screening Test result by pressing the "Manual" button while the subject is
still blowing.
6. Select
to start another test.
NOTE: A Screening Test only checks for the presence of alcohol in a breath sample. An accurate test result can
only be obtained via a Standard Test using a mouthpiece.
NOTE: Test records of Screening Tests cannot be printed or stored.
STANDARD TEST WITH INPUT
1. Attach a mouthpiece onto the sample port and select the Standard Test button.
2. The breathalyser will take an Air Blank Test to ensure the mouthpiece and sensor are clear of alcohol from
previous tests.
3. If the Air Blank Test result shows 0.000, the unit will proceed to test mode (skip the step 5).
4. If alcohol is detected, verify that a clean mouthpiece is being used and select "RETEST".
5. Use the stylus pen to input information. NOTE: Maximum 16 characters for each input.
Use
or
to move to the previous or next field. If no data fields are selected, the unit will proceed directly
to test mode. (Contact Andatech to enable/disable data fields)
6. When the screen displays "Please Blow", take a deep breath and blow into the mouthpiece with a slow and
steady breath until the BEEP tone stops.
7. The test result will display on the screen.
8. Select
to print the test result.
9. Select
to take another test.
10. If the subject does not blow into the unit within 30 seconds, the operator can either select "Refuse" to confirm
the subject refused to take the test, or "Test Again" to allow the subject to retry.
11. If the subject stops blowing too early or blows with insufficient force, select "Discontinue" to abandon the test,
or "Test Again" to retry.
12. The operator can also get an immediate Screening Test result by pressing the "Manual" button while the
subject is still blowing.
13. Push the button underneath the mouthpiece to remove the mouthpiece and take another test.
6