Andatech SOBERMATE S3 Panduan Pengguna - Halaman 4
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan Uji Andatech SOBERMATE S3. Andatech SOBERMATE S3 10 halaman.
WARNING DISPLAY
Full battery capacity
Empty battery capacity. Replace with brand new alkaline AAA batteries.
Testing cannot be performed because the battery is low.
Replace with brand new alkaline AAA batteries.
The breath sample was not strong enough for a measurement. Please
press the power button to take another test. Take a deep breath and then
blow into the device with a slow and steady breath. Only stop blowing when
you hear the double beep again.
If there has been no breath sample blown within 3 minutes, the device will
display this message. Press the power button to restart the countdown cycle
and retest.
The device is out of operational range due to the temperature. Bring the
device to operational temperature (according to the specifications) and
then perform a retest.
Excessive saliva may have entered the sensor during the test. Wait
20 minutes, then try again. If the error persists, please contact us.
The total number of tests has reached 500. Please send the device back to us
for calibration
SPECIFICATIONS
Indication of B.A.C
Accuracy
Warm up time
Sensor type
Response Time
Mouthpiece
Power supply
Battery life
Dimensions(WxHxD)
Weight
Calibration period
Operating temperature
0.000 - 0.400 %BAC
0.00 - 4.00 ‰
0.00 - 2.00 mg/L
±0.0075 %BAC at 0.100 %BAC
±0.075 ‰ at 1.0 ‰ at 25˚C
±0.0375 mg/L at 0.50 mg/L at 25˚C
15 - 20 seconds
*Warm up times may vary depending on the BAC measured
Fuel Cell Sensor
3 - 15 seconds
Sobermate brand mouthpieces
2 x AAA Alkaline Batteries
Approx. 1000 tests
51 mm x 109 mm x 18 mm
84 g (including batteries)
Every 12 months or 500 tests, whichever comes first
5 - 40°C