See above for readings from Palintest. 25 x dpd tests were taken. The average of which are
displayed. The difference between the 'very experienced' and 'less experienced' user is that the
experienced user would calibrate the colorimeter prior to calibration, take 3 tests and average
them out. The less experienced would simply take the dpd and compare. This is usually what
would happen in the vast majority of cases.
See below for typical coating of 'high manganese' application. This sensor still worked within
spec with this coating on the membrane. No need to calibrate OR rebuild. Just leave alone!
Q45H Training Handout 4.05
THIS DATA WAS PRODUCED BY PALINTEST