Advanced Measurement Technology ORTEC A-576 A-PAD Panduan Pengoperasian dan Servis - Halaman 11

Jelajahi secara online atau unduh pdf Panduan Pengoperasian dan Servis untuk Penguat Advanced Measurement Technology ORTEC A-576 A-PAD. Advanced Measurement Technology ORTEC A-576 A-PAD 11 halaman. Preamplifier, amplifier, discriminator with bias supply

Advanced Measurement Technology ORTEC A-576 A-PAD Panduan Pengoperasian dan Servis
&
6.7 on the 4
7 MeV range, or a factor of 4 on the 3
MeV range. The output of the biased amplifier is a
positive pulse with an amplitude, in the range from 0.1
to 10 V, which is proportional to the amount by which
&
8

Fig. 5. Detailed Block Diagram of the Model A-576 A-PAD.

7
the detected input energy exceeds the selected bias
level. The full-scale output of 10 V can be adjusted
down to 7.75 V, using the front-panel screwdriver
control (E).