Andon KD-5904 Panduan Pengoperasian - Halaman 4

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Monitor Tekanan Darah Andon KD-5904. Andon KD-5904 19 halaman.

DN:KD-5904-SMSY01 V3.0

SPECIFICATIONS

1. Product name: Blood Pressure Monitor
2. Model: KD-5904
3. Classification: Internally powered, Type BF applied part,IPX0,No AP or APG,Continuous
operation
4. Machine size: 145mm×104mm×55mm
5. Cuff circumference: 22cm-30cm, 30cm-42cm(Optional), 42cm-48cm(Optional)
6. Weight: about 300g (exclude batteries)
7. Measuring method: Oscillometric method, automatic inflation and measurement
8. Memory volume: 60 times with time and date stamp
9. Power source: DC:6V
10. Measurement range:
Cuff pressure: 0-300mmHg
Systolic:
Diastolic:
Pulse rate:
11. Accuracy:
Pressure:
Pulse rate:
12. Environmental temperature for operation: 5℃~40℃(41℉~104℉)
13. Environmental humidity for operation: ≤90%RH
14. Environmental temperature for storage and transport: -20℃~55℃(-4℉~131℉)
15. Environmental humidity for storage and transport: ≤90%RH
16. Environmental pressure: 80KPa-105KPa
17. Battery life: Approx. 360 times
18. All components belonging to the pressure measuring system, including accessories:
Pump, Valve, LCD, Cuff and Sensor.
Note: These specifications are subject to change without notice.

NOTICE

1. Read all of the information in the operation guide and any other literature in the box before
operating the unit.
2. Stay quiet, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same arm for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood
circulation in your arm to recover. Prolonged over-inflation (
or maintained above15 mmHg for longer than 3 minutes
of your arm.
7. Consult your physician if you have any doubt about below cases:
600mA, batteries: 4 ×1.5V
60-260mmHg
40-199mmHg
40-180 beats/minute
±3mmHg
±5%
SIZE AA
cuff pressure exceed 300 mmHg
) of the bladder may cause ecchymoma
4/19
Operation Guide
JS-KD5904-03-V2.0