Andonstar ADSM302 Panduan Pengguna - Halaman 18
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Mikroskop Andonstar ADSM302. Andonstar ADSM302 19 halaman.
FAQ
1. Why there says "missing files" during installation or initiator procedure?
Please confirm that Microsoft .NET Framework 4.0 and Microsoft Visual C++ 2010 Runtimes are installed
2. The software works fine, but the software can not recognize the microscope.
First, please make sure that you connect the camera and the PC properly and choose "PC camera" on the 5
inch monitor. Second, if there is still no image, please check the "Device Manager" of your PC, delete the "USB
Camera" in the "Imaging devices" and rescan . Let the system identifies again. After all, in the "Imaging devices"
there will show another name, which is "Andonstar Camera". Then restart the software.
3.Why there is a measurement error?
First, if you want to get a more precise result, you need to use a more precise calibration. Second, during the
whole process, make sure the magnification is always the same as the one when you set the calibration. Since
the camera is continuously magnifying, you'd better keep the same object distance also.
- 16 -