1. Connect the air input hose to a plant air source. Set plant air supply within 80 to 100 psi.
2. Attach the air input hose coupling to the dispenser.
3. Plug in the polarized foot pedal connector. Connect the foot pedal to the back of the dispenser
4. Check the voltage label on the input voltage selector cartridge. Connect the power cord into the back
of the dispenser. Connect the power cord into your local power source. (AC110V-220V / 50Hz)
5. Attach the 30cc barrel refilled with test fluid, maximum 2/3 full.
6. Use the correct Drifton piston to insert the barrel. Push the piston down into the barrel.
7. Remove the end cap, connect to the adapter assembly, twist clockwise to lock.
8. Push the black male quick-connector on the syringe barrel adapter assembly into the front of the
dispenser.
9. Remove the tip cap and replace it with an appropriate Drifton tip cap. Place the syringe barrel in the
barrel holder.
10. Initial setup is now complete. At this point you are now ready to set up your dispensing flow rate and
time to suit your application needs.
11. Note: Always use the lowest possible pressure and the largest possible tip size. The combination of
the lowest possible output pressure + largest possible tip size + longest possible dispense duration =
most consistent and accurate deposits.
12. During the initial testing, you will not use the vacuum control. Keep this control shut off (turned
completely clockwise - do not force)

Dispenser Hookup

5