ATLAS BIOTECNOLOGIES AB100 Buku Petunjuk - Halaman 7
Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Peralatan Laboratorium ATLAS BIOTECNOLOGIES AB100. ATLAS BIOTECNOLOGIES AB100 18 halaman.
and the preset volume) exceeds the permissible value
given in the table in section 1, the pipette recalibration
procedure should be carried out.
Before starting the recalibration it is necessary to check
whether the following requirements have been fulfilled
during error determination:
• the ambient temperature, and the temperature of the
pipette, tips and water was identical
• the density of the liquid used is close to that of distilled
water
• the balance with appropriate sensitivity has been used.
Volume checked [μl]
0,1 - 10
10 - 100
> 100
• mg/μl conversion factor has been taken into ac count
• the requirements given in sections 3 and 7 have been
fulfilled
If the above conditions are satisfied and the ac cu racy error
for selected volume given in section 1 exceeds the per-
missible value, the pipette recalibration procedure
should be carried out.
The recalibration can be performed within one full turn
of the key to the right or to the left only.
Recalibration conditions:
• Ambient temperature and the temperature of the
pipette, tips and liquid should be within the range 20-
25°C and stabilised during weighing within ±0.5°C
• Measurements should be conducted using distilled
water
• Balance sensitivity should be suitable for the vo lume to
be controlled
Recalibration procedure:
• Set the dose volume depending on the pipette volume
according to the following table:
9
Balance sensitivity [mg]
≤ 0,001
≤ 0,01
≤ 0,1
Range
Preset
Model
of the pipette
volume
volumes [μl]
[μl]
AB2
0.1 - 2
0.2
AB10
0.5 - 10
0.5
AB20
2 - 20
2
AB50
5 - 50
5
AB100
10 - 100
10
AB200
20 - 200
20
AB250
50 - 250
50
AB1000
100 - 1000
100
AB5000
500 - 5000
500
AB10000
1000 - 10000
1000
• Perform 5 aspirations, weigh each one and calculate
the average value of the aspirations.
• Calculate average aspirated volume in [μl] multiplying
the average aspiration amount [mg] by the distilled water
density coefficient [μl/mg], which depends on tempera-
ture and pressure according to the following table:
Temperature [°C]
95.0
20
1.0028
21
1.0030
22
1.0032
23
1.0034
24
1.0037
25
1.0039
If the average aspirated volume exceeds the per missible
value, the following should be done:
• Remove the pipetting pushbutton, (Fig. 4A),
Warning: The pipetting pushbutton consists of 2 parts:
the knob (Fig. 1A2) and the pushbutton (Fig. 1A1). After
removal of the pushbutton, both parts are separated.
• Holding the volume setting knob to protect it against
rotation, insert the calibration key into the cuts of the
calibration screw, (Fig. 4B),
•Turn the key clockwise to reduce the aspirated volume,
or counter-clockwise to increase the volume, (Fig. 4C).
ENGLISH
Volume change
ΔV for full turn
Permissible
of the
volumes
calibration key
[μl]
[μl]
(24 increments)
0.176 - 0.224
0.06
0.48 - 0.52
0.33
1.94 - 2.06
0.63
4.875 - 5.125
2.50
9.84 - 10.16
2.50
19.76 - 20.24
6.30
49.5 - 50.5
6.30
98.4 - 101.6
25.00
494 - 506
125.00
975 - 1025
250.00
Pressure [kPa]
101.3
105.0
1.0029
1.0029
1.0031
1.0031
1.0033
1.0033
1.0035
1.0036
1.0038
1.0038
1.0040
1.0040
10