Droper VET Panduan Petunjuk Pengoperasian - Halaman 11

Jelajahi secara online atau unduh pdf Panduan Petunjuk Pengoperasian untuk Produk Perawatan Hewan Peliharaan Droper VET. Droper VET 12 halaman.

KETONES
These aggressive agents may damage the plastic parts and cause the device to
malfunction.
ALCOHOL-BASED SPRAYS (20% - 40% alcohol content)
These cause the synthetic materials to tarnish and splinter.
IODINE SOLUTIONS
These may irreversibly stain certain clear plastic parts.
4.3
Re-installing the components
Once the parts are completely dry, re-insert the two components in the opposite
manner to the disassembly method, presenting each part opposite its tightening
clips. The plate must be presented with the infusion bag fastening hooks facing
the rear of the device. Visually check the correct installation of the different
elements.

5. PERIODIC INSPECTION

The device is designed to operate more than 6,000 times under normal operating conditions. However,
the manufacturer advises that users perform annual maintenance and inspections to ensure the
correct operation of the device, either by sending it to the manufacturer or distributor, or by using the
method described below :
Place a 1000 ml bag of NaCl 0.9%, the design of which corresponds to the plate's surface (e.g. : 1000 ml
Viaflo bag of NaCl 0.9% from Baxter) in the Droper Vet and introduce the spike of a classic infusion line
without a catheter or needle, and with the valve in the OFF position. Open the infusion line into a
recipient placed at the same height as the device and measure the time required to drain the bag. The
bag must be drained in less than or equal to 10 minutes.

6. STORAGE

The device must be stored in a temperature-controlled, dry place.
Relative humidity : 20% to 90%
Storage temperature : - 10°C to + 50°C
11/12