Clean‑Trace Luminometer LX25 Design and Components
1 . Power On/Off button
2 . Cursor control buttons
3 . Touch screen
4 . USB/Power port
5 . ATP Test Chamber
6 . Base
7 . Kick Stand
8 . Power cord and adapters
Battery Usage and Maintenance
The Clean-Trace Luminometer LX25 has an internal rechargeable
lithium ion battery .
Charging the battery
• Select and attach the charger adapter to the
power cord that is appropriate for power in
your country .
• Connect the power cord to the Clean-Trace
Luminometer LX25 .
• Connect the power cord to a power outlet .
• The battery becomes fully charged in
approximately 3 hours using the power supply .
If charging via PC connection, time to fully charge is
approximately 5 hours .
• To monitor the charge status, turn on the Clean-Trace
Luminometer LX25 . A lightning bolt appears in the battery
level indicator icon (Figure 1) .
When the battery is charged
• Disconnect the power cord from the
power outlet .
• Unplug the power cord from the Clean-Trace
Luminometer LX25 .
Using the battery level indicator
• The battery level indicator is located in the upper right
corner of the screen .
• When a Clean-Trace Luminometer LX25 is on, the indicator
shows the remaining battery level .
• In Figure 2, the indicator shows the battery is fully charged
• As the battery charge decreases, the blue segment of the
indicator becomes smaller .
• When the battery level is less than 10%, the indicator
becomes red, signifying the battery should be recharged .
Managing the battery
• During typical use the battery provides up to eight hours of
charge capacity .
• As the battery ages, it will provide less than eight hours of
charge capacity .
• The battery must be charged when the indicator is red,
indicating less than 10% charge capacity
• The battery does not need to be fully discharged
before recharging .
• When storing a Clean-Trace Luminometer LX25 for
more than a month, the battery level should be at
50% or less charge . Do not store a Clean-Trace
Luminometer LX25 with a fully discharged battery
because the battery may lose the ability to be recharged .
• Battery replacement is recommended after two years
of normal use or if the battery has been misused, or if
the Clean-Trace Luminometer LX25 is overheating or its
shape is distorted . Battery replacement is performed by
3M Health Care Service Center or authorized 3M Service
Provider . See Care and Maintenance .
Power On/Off/Sleep Mode
• To power up the Clean-Trace Luminometer LX25,
press and hold the grey power button for 2 seconds .
• To power off the Clean-Trace Luminometer LX25,
press and hold the grey power button for 2 seconds .
Accept prompts to shut down Clean-Trace
Luminometer LX25 .
• You can force the Clean-Trace Luminometer LX25 to
power off by holding the grey power button down for
20 seconds .
• The Clean-Trace Luminometer LX25 will automatically
power itself off after 2 hours of inactivity or if the battery
charge is 5% of remaining life . It is recommended to keep
the device charging when not in use .
• The Clean-Trace Luminometer LX25 will go into 'Sleep
Mode' after 1-30 minutes of inactivity .
Figure 1
To customize the time until sleep mode starts go to the
o
'Screen Timeout' screen which is found in the
Settings menu .
To exit sleep mode, press the "Up" cursor
o
control button.
• Sleep Mode vs . Power Off: The screen is blank for Sleep
Mode and when the device is Powered Off .
When you encounter a device with a blank screen, first
o
press the Up button, this will bring it out of Sleep Mode .
If the device does not respond to this, it is likely
o
Powered Off . Hold the Power Button for 2 seconds to
power on the device .