Drucker Diagnostics 614M Panduan Operator - Halaman 5

Jelajahi secara online atau unduh pdf Panduan Operator untuk Peralatan Laboratorium Drucker Diagnostics 614M. Drucker Diagnostics 614M 12 halaman. 12vdc mobile labaratory centrifuge

Drucker Diagnostics 614M Panduan Operator

Operation:

NOTE: Follow the initial setup procedure on page 3 before initial operation.
a) If any problems are found during the initial setup procedure, refer to the
troubleshooting section.
b) Plug the power cord into the automobile's 12V accessory outlet.
c) Turn the latch counterclockwise and open the lid.
d) Spin the rotor by hand; check for smooth rotation free from excessive wobble.
e) Place six test tube holders (supplied) of the same type inside the rotor (as
shown to the right), and verify that they are seated properly.
f)
Close the lid. Rotate the lid knob clockwise to its stop position.
g) Start a centrifugation cycle by rotating the timer knob.
h) Listen to the sound of the centrifuge. A smooth whirring sound should be heard.
Page 5