3M The Edge Panduan Memulai Cepat

Jelajahi secara online atau unduh pdf Panduan Memulai Cepat untuk Alat Ukur 3M The Edge. 3M The Edge 2 halaman.

3M The Edge Panduan Memulai Cepat
Turning on
1. To turn on the Edge, press the on/off
enter
The start-up screen will appear
indicating the model and firmware
version. Then, the measurement
view will display.
1
Resetting
2
Resetting the dosimeter is performed
using QSP-II software via the clear tab
3
The Edge may also be reset using the
keypad. (See Chapter 3 and Chapter 4 in
the Edge user manual for details.)
4
Navigating
. Press the up/down arrow key to
5
1
6
navigate through measurement values,
7
battery condition, and run time.
Setting-up the Edge
1. Windscreen & Microphone
All the Edge parameters are set up via
QSP-II software.
2. LED indicator
Chapter 3 in the Edge user manual for
Red – indicates dose has
details).
exceeded the alarm level.
Amber – indicates
stopped.
Calibrating
Green – indicates
1. Remove the windscreen.
measuring.
2. Ensure The Edge is turned on and
3. Measurement display
stopped.
3. Slide the Edge into the mouth of the
4. Up arrow
calibrator (ensuring the adapter ring is
inserted into the calibrator).
5. Run/Stop key
4. Turn on the calibrator. Ensure the
calibrator is set to 114 dB and 1000 Hz.
6. On/Off Enter key
See subsequent steps on backside.
7. Down arrow
key.
from the Edge Set-up button.
(See