DS SG6000 Panduan Pengguna - Halaman 3

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Generator Portabel DS SG6000. DS SG6000 4 halaman. Rf signal generator

DS SG6000 Panduan Pengguna
11. Fine Power Tune – Adjust output power in fine (~0.05dB) increments between the 0.5dB steps.
12. Message Bar – Displays feedback information sent from the device
Control Software Notes
If reference auto-detection is misbehaving with your input signal, it can be forced to external or internal using
the select box and "Change Ref" button. A SG reset is recommended after this setting is changed.
"Send Command" allows a user to send a raw SCPI command to the Signal Generator, in USB mode the GUI will
also display any response from the device in the message bar at the bottom.
The Vernier setting allows fine power output level setting between the 0.5dB steps. Highest power output
results when the slider is farthest to the right.
If the USB voltage box goes into the red, performance will be unpredictable or an error will be indicated by the
device. Try finding a better USB power source or cable.
If the device encounters any errors it will start blinking red, to query what the error is, send the SCPI command
"SYST:ERR?".
Hardware
The MCX 10MHz reference input will accept sine or square wave input between -10 and +10dBm.
The sweep trigger MCX input is pulled high to +3.3V internally, when pulled low the saved sweep will be
initiated. Sweep parameters must be loaded from the GUI for this trigger to be activated. This input is 5-volt
tolerant. Brining the input back high will not stop the sweep or have any other effect.
Sweeping Details
The Sweep mode control lets the user configure a
continuous or single sweep.
Direction UP increases frequency from Start to Stop,
DOWN starts the sweep at the stop frequency.
The Trigger option changes the functionality of the
physical trigger input pin between activating the
complete sweep, or just the next STEP. This is only
available in certain firmware versions.
The Points field determines how many independent
frequency points are calculated total. This can be
from 2 to 10000.
The Dwell time field controls how long the
synthesizer will pause at each new frequency point.
Minimum value is ~3ms.
The Start and Stop frequency settings control the
beginning and end of the sweep.
The Step Size and Run Time are system calculated values based on the above values. Run time is only a close
estimate of the ideal value, not including actual minor system overhead and phase lock loop delays.
The Save Sweep button transfers these values from the GUI to the signal generator CPU. This button is critical to
press after values are updated in the control software.
The Abort button immediately stops the device mid sweep.
The Trigger button emulates the physical trigger input pin by sending a command to start the sweep now.
Version 4.0 – DS Instruments 2022 – www.dsinstruments.com