Anritsu Company 34NN50A Panduan Kalibrasi - Halaman 13

Jelajahi secara online atau unduh pdf Panduan Kalibrasi untuk Adaptor Anritsu Company 34NN50A. Anritsu Company 34NN50A 16 halaman. Precision coaxial adapters

Performance Verification
9. Under the Set Scale Key:
a. Set the Resolution to 500 mU/div.
b. Set the Reference Value equal to 1.1
10. Install the Adapter Under Test to VNA Port 1.
11. Install the appropriate termination from the 3653 cal kit to the
open end of the Adapter Under Test.
12. Verify that the SWR of the Adapter Under Test is within
specification (refer to
Hint: Use a marker to find the maximum SWR value.
Model 34 Series CG
Table 2-1, "Model 34 Series
PN: 10100-00057 Rev. B
2-3 Test Procedure
Specifications).
2-3