3M W-2806/07006 Petunjuk Pengguna - Halaman 13
Jelajahi secara online atau unduh pdf Petunjuk Pengguna untuk Monitor 3M W-2806/07006. 3M W-2806/07006 19 halaman. Compressed air filter and regulator panel
Juga untuk 3M W-2806/07006: Pengoperasian Umum dan Petunjuk Suku Cadang (2 halaman)
Calibration
Calibrate the monitor in the temperature range in which it will be used. After each adjustment
in the following steps, allow time for the changes to stabilize. Recheck all adjustments. Turn
monitor on and allow at least 5 minutes warm-up before calibration and adjustments are made.
Zero Adjustment
Remove sample tube from the black tube lock collar on top of the flow meter. Introduce a sample
of zero air to the sample tube (see User Instructions for the 3M
TM
Calibration Kit 529-04-48 or
529-04-49). The sample flow should be between 0.5 and 1.5 scfh.
Verify that the green NORMAL light is on, the red ALARM light is flickering about once a second
and the display is showing "00". If the display is showing other than "00" , adjust the ZERO
potentiometer (next to the display) so that the reading is "00" . Clockwise adjustment increases
reading. Notice that the display jumps from 00 to +02 or to –02 (monitor does not display –01 or
+01). Try to set the potentiometer midway between the two extremes. Remove the zero air sample.
Span Adjustment
Introduce a known sample of 50 to 150 ppm CO to the sample tube (see User Instructions for the
3M
TM
Calibration Kit 529-04-48 or 529-04-49). The sample flow should be between 0.5 to 1.5 scfh.
Verify that the green NORMAL light is off, the red ALARM light is on full bright and the
stabilized reading, after approximately one minute, is equal to the known concentration of CO.
If the display is showing a difference, adjust the SPAN potentiometer (next to the display) to
obtain proper value. Turn clockwise to increase reading. If the span adjustment cannot be made
as indicated, the sensor needs to be replaced; make sure this is not happening due to an empty
span gas cylinder. See Replacement Part section of this User Instruction.
Remove the known sample of CO and reconnect the sample tube to the black tube lock
collar on top of the flow meter. The instrument is now properly adjusted for use.
12