4.1 Environmental and transport conditions: 
 
Internal use.  Altitude lower than 3000 m (10,000 ft). 
Temperature: 18°C / 40°C (64°F / 104°F). 
Relative Humidity: < 80% (non‐condensing). 
 
4.2 Safety precautions: 
 
a)
Do not insert anything inside the motor. 
b)
The "Motor" is supplied non‐Sterile. 
c)
Use only power cord as described in the section Package Contents 
d)
Use for intended purposes only. Failure to observe the operating instructions may result in the patient or user suffering 
serious injury. Before using this product, make sure that you have studied and understood the operating instructions. 
e)
Do  not  install  where  there  is  a  risk  of  an  explosion.  The  Systems  are  not  intended  for  operation  in  the  presence  of 
flammable anesthetics or gases. 
f)
Do not disassemble or alter the System motor, console, or foot switch. 
g)
Connect mains power cable to a properly grounded outlet only. 
h)
Never touch drills, burs, or other handpiece tips when they are still rotating. 
i)
Handpiece should only be attached when the motor has stopped running. 
 
4.3 Setting up the unit: 
 
1.
Cut the adhesive tape and open the suitcase.  Remove the unit and its accessories from their original packaging and place 
on a flat surface. Not install near heat sources, direct or indirect. 
2.
Verify the contents of the box according to the section "PACKAGE CONTENTS" 
3.
Do not kink the motor hose since it may damage it. 
4.
All accessories are supplied non‐sterile except for tubing. 
5.
The serial number on the Unit's rear panel must be the same as the one on transport documents. 
6.
Allow at least 150 mm free space around the unit for cooling ventilation. 
 
Damage in Transit 
If the packaging is visibly damaged on delivery, please proceed as follows: 
a.
The recipient of the package must record the loss or damage on the delivery receipt. The recipient and the representative of 
the shipping company must sign this delivery receipt. Without this evidence, the recipient will not be able to assert a claim for 
damages against the shipping company. 
b.
Leave the product and packaging in the condition in which you received it and do not use the product. 
If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows: 
a.
Report the damage immediately or at least 7 days after the delivery to the delivery company. 
b.
Leave the product and packaging in the condition in which you received it and do not use a damaged product. 
 
4.4Assembling the unit 
 
Micro Motor connection 
Connect the micro motor into the receptacle on the front of the central unit.      
Connect the Contrangle to the Micro Motor. 
Only for "REF" ABNE300 press the button on the rear side to switch on LED Light 
Place the Micro Motor on the motor holder. 
Foot Pedal connection 
Connect the foot pedal into the receptacle on the front of the central unit.  
Place the foot pedal on the ground. 
 
While inserting the plug, align the guide on the connector in an upright position to fit into the corresponding female guide on the B‐ONE. 
Connectors have a spring retention system to help avoid accidentally disconnecting the unit's cords and cables. 
 
 
 
 
 
 
USER MANUAL Rev 25/05/2023 
 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
B‐ONE 
Page 6 of 16 
4. SET UP