DST ObservAir Panduan Pengoperasian - Halaman 5

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Aksesori DST ObservAir. DST ObservAir 20 halaman. Distributed sensing technologies

proprietary environmental training approach. Prior to delivery, all ObservAir
units sample clean ('zero') air for at least 24 hours while being subjected
to fluctuating environmental conditions. Using the data collected during this
training period, the unique environmental dependence of each ObservAir
unit is modeled mathematically. The models are uploaded to each unit, and
used to correct air pollution concentration measurements in real-time. Each
ObservAir is delivered with its own unique zero-calibration sheet, as shown
in Figure 3 below. The calibration sheets show the sensor's baseline
environmental dependence and measurement performance both before
and after compensation. The sensor also carries out regular calibration and
diagnostic checks of the underlying electronics, and includes other software
features to maintain and validate sensor performance.
Figure 3. Each ObservAir comes with its own zero-calibration sheet

1.3. Base package contents

Each ObservAir comes with the following base set of accessories and
supplies:
Magnetic cover
10 replacement filter tabs
Micro-SD card with 16 GB capacity
Charger (US Plug) and 3-foot (1 m) micro-USB cable
3-foot (1 m) length of conductive sample line
Zero-calibration sheet (hardcopy)
Quick-start guide (hardcopy)
8
ObservAir Operating Manual

2. Technical Specifications

The ObservAir's technical specifications are summarized in the tables
below. All electrochemical cells are sourced from Spec Sensors. Gas
measurement performance specifications are adapted from data provided
by the manufacturer, and where possible, DST validation of the
electrochemical cells in the ObservAir platform. All baseline measurement
noise specifications are derived from data collected while the ObservAir is
sampling clean ('zero') air. More information on the electrochemical cells
can be found at www.spec-sensors.com.
All measurement performance specifications are derived from ObservAir
data collected at a sample flow rate of 100 ccm near standard
atmospheric conditions: Temperature and relative humidity (RH) ranging
from 15 to 30ºC and 25 to 40%, respectively.

2.1. General specifications

Standard: Black carbon (BC) aerosol
Air pollution
measurement species
Optional: CO, NO
Black carbon: Filter-based light absorption (880 nm)
Principle of operation
Gases: Electrochemical cells
Standard: Wi-Fi, Bluetooth, USB
Communications
Optional: LTE, LoRa, SigFox (choose one)
Sample air flow rate
50 to 200 ccm
Sample interval
2 to 60 seconds
1.2 W (at 100 ccm flow rate)
Power consumption
Battery life
≥ 24 hours (at 100 ccm flow rate)
Flow rate (ccm)
Filter life
(BC
= 1μg/m
3
)
avg
Filter life (days)
Data storage
Removable SD card (16Gb card provided)
Operating conditions
Temperature: 5 to 40ºC; RH: 15 to 80%
Dimensions/Weight
120 x 80 x 45 mm / 600 grams
Charging
5V DC at 2.1A max (microUSB and charger provided)
Table 2. ObservAir general specifications
, SO
, H
S, O
, ethanol (up to 2)
2
2
2
3
50
125
200
6.3
2.5
1.6
ObservAir
Operating Manual
9
®