5.4 Electrical Safety
Ensure a grounded/earthed connection
WARNING
To avoid the risk of electric shock, the MOVE/MOVE+ cart
must only be connected to a supply mains with grounding/
protective earth . The MOVE/MOVE+ cart is equipped with
a three-pin (US: prongs) grounding/earth type plug . If you
cannot insert the plug into your outlet, contact your elec-
trician to replace the plug or outlet, to ensure a grounded
power connection . Do not try to defeat or ignore the
purpose of the grounding-type plug .
Electrical shock
WARNING
There is a risk of electrical shock if you attempt to access
the inside of any part of the system . Only authorized and
qualified service personnel may access the inside of any
part of the system .
Do not touch the electrical terminals
WARNING
The smart cord, the scanner tip, and the scanner tip inter-
face surface on the scanner contain electrical terminals .
For electrical safety, and to prevent corrosion and other
damages, the electrical terminals must not be touched at
any time, including touching electrical contacts and the
patient at the same time .
Do not use a power strip
WARNING
Do not connect the MOVE/MOVE+ cart to an ordinary power
strip or extension cord, because these may not be secure in
relation to grounding . If the ground connection fails, these
can cause the following hazards and potential injury:
- The total leakage current for all the connected equipment
can exceed the limits specified in EN (IEC) 60601-1 .
- The impedance of the ground connection could exceed
the limits specified in EN (IEC) 60601-1 .
Stress on cables
CAUTION
All externally connected cables (the power cable,
network cable, and the smart cord) must never
be subjected to pulling stress .
Spilled liquids
WARNING
Do not bring liquids, such as beverages, near the TRIOS IOS
system . Do not spill liquids on the TRIOS IOS system .
Condensation
WARNING
Variations in temperature or humidity may cause water
to condense inside the system, which can cause damage .
Always let the system reach room temperature before you
plug it into the power source .
If the system has been subject to major temperature or
humidity changes, wait till the system has reached room
temperature before plugging it into the power source .
If there is visible evidence of condensation, wait at least 8
hours before plugging it into the power source, to ensure
that no internal condensation occurs .
Disconnecting from mains
WARNING
There is no power ON/OFF switch on the system; therefore,
the only reliable means to disconnect the system from
mains is to unplug the power cord . Do not position the
system so that it is difficult to unplug the power cord .
Please allow enough time for the system to shut down
properly before disconnecting the power cord .
System S3P (pod) only: You must also disconnect
the PC from mains .
PC connection options and requirements
System S3P (pod) only
WARNING
The PC is a standard IT equipment certified to IEC 60950-1
and is located at least 1 .5 m away from the patient . If you
choose this setup, you must not scan the patient and touch
the PC at the same time! You must not touch any periph-
erals connected with a cable to the PC while scanning a
patient . See the PC's instructions for use for further details .
USB connection options and requirements
System S3A (MOVE/MOVE+ cart) only
WARNING
Only bus-powered USB devices without external power
supply that support USB Full speed mode must be con-
nected to the USB port .
9