3Shape TRIOS S3A-18 Setup Manual - Halaman 8

Jelajahi secara online atau unduh pdf Setup Manual untuk Peralatan gigi 3Shape TRIOS S3A-18. 3Shape TRIOS S3A-18 14 halaman.

3Shape TRIOS S3A-18 Setup Manual
In case of equipment failure
WARNING
If at any time the TRIOS IOS system malfunctions,
or if you suspect in any way that the TRIOS IOS system
is not working correctly:
• Remove the scanner from contact with the patient .
• Unplug the system and make sure it cannot be used
before it is checked .
• Contact your reseller .
• Do not attempt to open any covers on the TRIOS IOS system .
Portable RF communications equipment
WARNING
Portable RF communications equipment (including periph-
erals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the
MOVE/MOVE+ cart system, including cables specified by
the manufacturer . Otherwise, degradation of the perfor-
mance of this equipment could result .
Wireless Connection
The wireless scanner is intended to be used indoors, in a domestic
or commercial electromagnetic environment .
CAUTION
Other portable and mobile RF communication equipment
could interfere with the wireless scanner, even if the other
equipment complies with CISPR emission requirements .
CAUTION
The wireless scanner could interfere with other portable
and mobile RF communication equipment .
CAUTION
It is recommended to keep a line of sight between the
wireless scanner and the USB wireless dongle attached
to the PC .
5.2 Mechanical Hazards
Fragile and movable parts
WARNING
Be careful to avoid the following potential sources of injury:
• Do not lean or sit on any part of the system .
• Movable parts can pinch body parts .
Dropped or damaged equipment
WARNING
If you drop a scanner tip on the floor, you MUST dispose
of it immediately and NOT use the same tip again for
scanning . There is risk that the mirror in the tip has become
dislodged and can fall out .
WARNING
If the handheld scanner is dropped or bumped, it should
immediately be calibrated before further use . If calibration
fails, please contact your technical service provider . See this
guide's instructions on calibrating the handheld scanner .
8
WARNING
If the handheld scanner is dropped, examine the TRIOS
IOS system for any mechanical damage . If a mechanical
damage is observed or suspected, treat this as an equip-
ment failure (see this guide's instructions on handling
equipment failures) .
WARNING
If a battery is dropped, examine it for any mechanical
damage . If a mechanical damage is observed or suspected,
the battery must be disposed of in the correct manner .
System S3A (MOVE/MOVE+ cart) only
WARNING
• Keep both hands on the upper part of the MOVE/MOVE+
cart when moving it across the floor .
• Do not attempt to lift the MOVE/MOVE+ cart alone . When
moving the system over a door step or other unsmooth sur-
face, get assistance from another person to ensure stability .
• When moving the MOVE/MOVE+ cart across the floor,
maintain all its wheels in contact with the floor and ensure
that they are not locked .
• Always place the scanner in its intended mount on the
MOVE/MOVE+ cart when not in use and when moving
the MOVE/MOVE+ cart .
• Keep your feet clear of the MOVE/MOVE+ cart's base
and wheels when moving the MOVE/MOVE+ cart .
• Never drag the MOVE/MOVE+ cart across any of its cables .
• Do not place the MOVE/MOVE+ cart on a slope greater
than 10 degrees .
• Keep sharp objects away from the touch screen .
System S3P (pod) only
WARNING
• Always place the scanner in its intended mount on the
pod when not in use .
• Do not place the pod on a slope greater than 5 degrees .
5.3 Explosion Hazards
Environment
WARNING
The TRIOS IOS system is not designed to be used in poten-
tially explosive environments, such as in close proximity
to flammable liquids or gases, or in oxygen-enriched
atmospheres .
Distance to the patient with anesthetic
WARNING
There is a potential explosion hazard if the TRIOS IOS
system is operated in the presence of flammable anesthetic .