Apogee S2-442 Panduan Pemilik - Halaman 17

Jelajahi secara online atau unduh pdf Panduan Pemilik untuk Aksesori Apogee S2-442. Apogee S2-442 19 halaman. Par-far sensor

Apogee S2-442 Panduan Pemilik

TROUBLESHOOTING AND CUSTOMER SUPPORT

Independent Verification of Functionality
Apogee S2-141 PAR-FAR sensors are self-powered devices and output two voltages, one proportional to PPFD and
one proportional to far-red photon flux density. A quick and easy check of sensor functionality can be determined
using a voltmeter with millivolt resolution. To check the PAR sensor, connect the positive lead wire from the
voltmeter to the yellow wire from the sensor and the negative (or common) lead wire from the voltmeter to the
blue wire from the sensor. Direct the sensor head toward a light source and verify the sensor provides a signal.
Increase and decrease the distance from the sensor head to the light source to verify that the signal changes
proportionally (decreasing signal with increasing distance and increasing signal with decreasing distance). Blocking
all radiation from the sensor should force the sensor signal to zero. To check the far-red sensor, repeat the process
above by connecting positive lead wire from the voltmeter to the white wire from the sensor and the negative lead
wire from the voltmeter to the black wire from the sensor.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
Apogee S2-141 PAR-FAR sensors are calibrated with approximate calibration factors of 100 mol m
PPFD measurement and 50 mol m
sensitivities of 0.01 mV per mol m
measurement device (e.g., datalogger or controller) should have voltage measurement resolution of at least 0.01
mV in order to provide PPFD resolution of 1 mol m
resolution of 1 mol m
-2
An example datalogger program for Campbell Scientific dataloggers can be found on the Apogee webpage
(https://www.apogeeinstruments.com/content/PAR-Far-Sensor-Unamplified.CR1).
Cable Length
When the sensor is connected to a measurement device with high input impedance, sensor output signals are not
changed by shortening the cable or splicing on additional cable in the field. Tests have shown that if the input
impedance of the measurements device is greater than 1 mega-ohm there is negligible effect on the calibration,
even after adding up to 100 m of cable. All Apogee sensors use shielded, twisted pair cable to minimize
electromagnetic interference. For best measurements, the shield wire must be connected to an earth ground. This
is particularly important when using the sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See Apogee webpage for details on how to extend sensor cable length
to-make-a-weatherproof-cable-splice/).
-2
-1
s
per mV for far-red photon flux density measurement, yielding approximate
and 0.02 mV per mol m
-2
-1
s
-1
s
.
-2
-1
s
, respectively. Thus, a compatible
-2
-1
s
and 0.02 mV in order to provide far-red photon density
(http://www.apogeeinstruments.com/how-
-2
-1
s
per mV for