Apogee SQ-521 Manual - Halaman 4

Jelajahi secara online atau unduh pdf Manual untuk Aksesori Apogee SQ-521. Apogee SQ-521 12 halaman.

Apogee SQ-521 Manual
USING APOGEE QUANTUM SENSORS WITH ZENTRA SYSTEM

1. INTRODUCTION

The SQ-521 Full-Spectrum Quantum sensors from Apogee Instruments, Inc. are high-
accuracy, single-band radiometers designed for continuous measurement of photosynthetic
photon flux density (PPFD) or photosynthetically active radiation (PAR) measurement in
indoor or outdoor environments. The Apogee Full-Spectrum Quantum sensor has nearly
equal sensitivity across the spectral range from 389–692 nm (PAR band is 400–700 nm)
and is therefore a good choice for both above- and below-canopy measurement in outdoor
environments and also for indoor environments, where artificial light sources are used.
The information in this document explains how to install the required hardware to mount
Apogee SQ-521 sensors that have been preconfigured by METER Group to work seamlessly
with METER ZENTRA series data loggers. Details of how the ZENTRA system handles the
data are also included. Please read this document carefully in its entirety before going out to
the field.
For more information on Apogee Full-Spectrum Quantum Sensor, please review the
SQ-521 user manual
spectrum-Quantum-sensor).

2. INSTALLATION

Follow the steps listed in
bracket, leveling plate, and nylon screws are included with the sensor. Other tools will need
to be provided.
Tools Needed
(apogeeinstruments.com/sq-521-ss-sdi-12-digital-output-full-
Table 1
to install Apogee sensors in the field. A cable, mounting
Table 1 Installation
Wrench 13 mm (0.5 in)
Flathead screwdriver
Mounting post 33.0 to 53.3 mm (1.3 to 2.1 in) diameter post, pole, tripod,
tower, or other similar infrastructure that extends above the canopy
Mounting bracket + leveling plate Model AL-120
Nylon screw 10-32 x 3/8
METER ZENTRA series data logger ZL6 or EM60
METER ZENTRA software ZENTRA Utility, ZENTRA Utility Mobile, or
ZENTRA Cloud
2