Spectral Response
Temperature Response
Mean spectral response measurements
of six replicate Apogee SQ-100 (original)
and SQ-500 (full-spectrum) series
quantum sensor. Spectral response
measurements were made at 10 nm
increments across a wavelength range
of 300 to 800 nm in a monochromator
with an attached electric light source.
Measured spectral data from each
quantum sensor were normalized by the
measured spectral response of the
monochromator/electric light
combination, which was measured with
a spectroradiometer.
Mean temperature response of ten SQ-500
series quantum sensors (errors bars
represent two standard deviations above
and below mean). Temperature response
measurements were made at 10 C intervals
across a temperature range of approximately
-20 to 50 C in a temperature controlled
chamber under a fixed, broad spectrum,
electric lamp. At each temperature set point,
a spectroradiometer was used to measure
light intensity from the lamp and all quantum
sensors were compared to the
spectroradiometer. The spectroradiometer
was mounted external to the temperature
control chamber and remained at room
temperature during the experiment.