OPERATION AND MEASUREMENT
Wire up to green block and plug in making sure to connect the ground wire first. Then, connect the signal line
second and the power line last. Below is common wiring up of Apogee Instruments SDI-12 sensors. Please check
your individual sensor manual for wiring instructions.
VERY IMPORTANT
: Apogee changed all wiring colors of our bare-lead sensors in March 2018 in conjunction
with the release of inline cable connectors on some sensors. To ensure proper connection to your data device,
please note your serial number or if your sensor has a stainless-steel connector 30 cm from the sensor head then
use the appropriate wiring configuration.
Common Apogee Sensor Wiring Configurations
Below are the two possible sensor wiring configurations of Apogee sensors. Please see your respective sensor's
product manual to find the correct wiring configuration for your sensor.
Red: Power In (4.5-24 V DC) (connect last)
Black: SDI-12 Data Line
Clear: Ground (shield wire) (connect first)
Black: Ground (for sensor signal and input
power) (connect first)
Red: Power In (4.5-24 V DC) (connect last)
White: SDI-12 Data Line
Clear: Shield/Ground