Apogee Instruments SU-300 UV-I Panduan Pemilik - Halaman 8
Jelajahi secara online atau unduh pdf Panduan Pemilik untuk Aksesori Apogee Instruments SU-300 UV-I. Apogee Instruments SU-300 UV-I 19 halaman. Ultraviolet index & ultraviolet-b sensor
Cosine Response
Cosine Response Graph
10
8
6
4
2
0
-2
-4
-6
-8
-10
0
10
20
30
40
50
60
Solar Zenith Angle [°]
Directional, or cosine, response is
defined as the measurement error at
a specific angle of radiation incidence.
Directional error for Apogee SU-300
series UV-I/UV-B sensors is
approximately ± 2 % and ± 5 % at solar
zenith angles of 45° and 75°,
respectively.
Mean cosine response of four
Apogee UV-I/UV-B sensors. Cosine
response was calculated as the
relative difference of UV-I/UV-B
sensors from a reference UV-I/UV-B
sensor deployed outdoors. These
data are the average of the AM and
PM response.
70
80
90