Instrument Calibration
The instrument must be calibrated before taking measurements. Calibration involves measuring a black
trap, a white tile and for gloss units, a black tile. Each instrument also comes with calibration stand.
Datacolor 45 Quick Start Guide v1.0 – 003‐400063
Select the calibration option by selecting the calibration icon
in the live bar.
The calibration screen is displayed. Notice that the calibration
indicator in the title bar is red indicating that the instrument
is not calibrated.
The last calibration time is shown as well as the time for the
next calibration.
To begin the calibration, either place the instrument on the
black trap on the tray or position the instrument and black
trap on the stand.
Press either of the measurement buttons on the instrument
to take the measurement.
After the black trap is measured, the program will ask for the
white tile to be measured.
To measure the white tile, either place the instrument on the
white tile on the tray or position the instrument and white
tile on the stand.
Press either of the measurement buttons on the instrument
to take the measurement.