TROUBLESHOOTING

PROBLEM
The mixer cannot be
powered on
Abnormal sounds
or vibrations during
operation
Display panel does
not light up
Unit operation has
delayed start
Mixing cup not closing
or opening well
Mixing results are poor
Display panel shows
'E1'
Display panel shows
'E2'
Display panel shows
'E3'
When requested to send the unit back for repair, please send the full unit and power cord for
further inspection.
CAUSE
Power cord is disconnected or
main power switch is off
The fuse is blown
Motor screws at the base of the
unit haven't been removed
Operation of the mixer without
mixing cup or with empty cup
Lid of mixing cup is loose
Defective gears
Defective CPU Board
Defective CPU Board
Mixing cup deformed
Blocked air hole in the lid
Incorrect mixing ratio of alginate
and water, or incorrect mixing
time
Incorrect operating steps
Poor or expired alginate
powder
Deteriorated rotation speed
The lid of the mixer isn't
completely closed
Defective microswitch
The lid buckle of the mixer does
not work
Defective microswitch
Unstable operating platform
Defective microswitch
SOLUTION
Plug the power cord in again and
check the main power switch
Inspect the fuse and change if
needed
Remove the motor fixing screws
as instructed in page 5
Stop operation
Stop operation and completely
close the lid
Return to the dealer for repair
Return to the dealer for repair
Return to the dealer for repair
Replace mixing cup
Free the air hole
Refer to instructions from
alginate supplier
Refer to Panel Operations
Section C on page 6
Use a new batch of alginate
powder
Return to the dealer for repair
Open and fully close the lid again
Return to the dealer for repair
Open and fully close the lid again
Return to the dealer for repair
Ensure operating platform is
stable
Return to the dealer for repair