Datalogic MAGELLAN 8400 Lembar Referensi Cepat - Halaman 2

Jelajahi secara online atau unduh pdf Lembar Referensi Cepat untuk Pemindai Datalogic MAGELLAN 8400. Datalogic MAGELLAN 8400 2 halaman.
Juga untuk Datalogic MAGELLAN 8400: Panduan Memulai Cepat (18 halaman)

Datalogic MAGELLAN 8400 Lembar Referensi Cepat
Lift Handle
9
Remove the platter and
push the wire lift handle
to the upright position in
preparation for sealing the
calibration door.
IF YOU ENCOUNTER PROBLEMS OR HAVE QUESTIONS ABOUT
October 2013
10
Fully close the calibration
door and apply the
calibration seal in accor-
dance with local require-
ments. The door is fully
closed when the holes in
the door and the metal
tab are aligned. (See
Sealing Notes.)
ANY PART OF THE CALIBRATION PROCESS:
CALL DATALOGIC TECHNICAL SUPPORT:
VIEW THE CALIBRATION VIDEO AT: http://www.youtube.com/watch?v=wHYjcTdd9Gk
1-800-547-2507
820061014 (Rev. A)
Sealing Notes
No portion of the seal,
whether wire, paper
label or other method,
should be in contact
with the weigh frame
(shown in green) or
scale movement will be
compomised.
Conditions to Avoid:
Wire loop too long (not snug)
Calibration door not closed
Seal in contact with weigh frame