Dymax ACCU-CAL™ 150 Radiometer User Guide
Operation
Figure 2. ACCU-CAL™ 150 Radiometer Face Plate
SETUP Mode Button
Soft Button Operation
Buttons perform the
function indicated on the
Display Window bottom
line.
Unit Serial Number
Battery Door
1. Turning ON the radiometer: Press and hold the ON/OFF button until the Window Display illuminates.
The Window Display will briefly show the radiometer model name, serial number, software version,
date of calibration, range, and wavelength band installed. Continuing to depress the ON/OFF button
at startup will display the startup information until the ON/OFF Button is released. The Display
Window will then enter the default mode and show the data from the last run before the unit was
turned off.
2. Turning OFF the radiometer: Press and hold the ON/OFF button. A tone will sound. When the tone
stops, release the button. The unit will turn off.
3. Entering RUN mode: A short press of the RUN/EXIT Button clears the memory and puts the unit in
RUN mode. The Display Window shows "RUNNING" after quickly displaying the internal temperature
of the unit. Confirm that the unit shows "RUNNING" before initiating a reading.
4. Place the radiometer on the belt or object with the Optic Window (Figure 3) looking toward the UV
source. The Display Window and buttons will be facing away from the UV source. When the
radiometer exits the curing chamber, the Display Window will still be flashing "RUNNING".
CAUTION:
Exposing the Display Window to high UV radiation will damage the display.
Display Window
ON/OFF Button
Depress and Hold until
screen is illuminated.
RUN/EXIT Button
Press to enter RUN
Mode.
Press to Exit RUN
Mode.
USB PC Connector
(side of unit)