Troubleshooting

Problem
The LED array will not
turn on and the "LED
message
INHIBIT OPEN"
is displayed
The LED array will not
turn on and the "LED
INHIBIT OPEN" message
is not displayed
The LED array will not
turn on and the error
screen is visible
NOTE: If your unit displays an error message on the LCD screen, refer to the troubleshooting section of the
BlueWave
AX-550 LED Curing System User Guide for more information.
®

Spare Parts and Accessories

Item
Cables
Interlock Cable
Radiometer
ACCU-CAL™ 160-LED
Misc. Replacement Parts
Work Surface Tray Assembly
|
12
D Y M A X . C O M
Possible Cause
The light shield's door not fully
closed
Interlock cable is not fully
plugged in on both ends
Reference the Dymax BlueWave
information.
Part Number
60471
41585
40806
Corrective Action
Close the door until it clicks
Firmly secure the interlock cable on
both ends. Verify the connector
screws are engaged and fastened
AX-550 User Guide for further
®