Troubleshooting

Table 1. Troubleshooting Chart for Model 200 Dispensing Valve
Problem
The valve does
not dispense
anything
The dispense
rate is too fast
The dispense rate
is too slow
There are air
bubbles in the
fluid
Material leaks
from the valve tip
Possible Cause
The fluid pressure is too low
There is material cured in the
Fluid Section
The Diaphragm is worn
The fluid pressure is set too
high
Stroke Adjustment Nut (E) set
too high
Dispense tip gauge too large
Fluid pressure set too low
Stroke Adjustment Nut (E) set
too low
Dispense tip gauge too small
Valve not properly purged
There is a problem with the
Fluid Reservoir and Delivery
System
The Diaphragm is worn
There are air bubbles trapped
in the fluid body or in the
dispense needle
Model 200 Dispensing Valve User Guide
Corrective Action
Increase the fluid pressure above
15 psi
Disassemble and clean the valve
Replace the Diaphragm (B)
Decrease the fluid pressure
Turn the Stroke Adjustment Nut
(E) counter-clockwise
Replace the Dispense Tip with a
smaller size tip
Increase the fluid pressure
Turn the Stroke Adjustment Nut
(E) clockwise
Replace the Dispense Tip with a
larger size tip
Point the valve up and cycle until
air bubbles are removed
Diagnose and repair
Replace the Diaphragm (B)
Point the valve up and cycle it
until the air bubbles are removed
13