Air escape holes are present in the base of the
pellet cup and should be checked monthly to
ensure that they remain open. Use a .036" or
thinner diameter wire to open plugged holes.
Also, make sure the rim of the pellet cup and
the four blocks that support the cup are clean.
Clean and lubricate top gasket every 3 months.
Apply a drop of liquid dish soap onto the
gasket on the pressure chamber with a
fingertip. Identify potential gasket cuts or
irregularities. Remove the soap from the gasket
with a soft cloth.
Lubricate the side ring once each year.
Remove the gasket ring from the chamber.
Using a light grease (016-021), lubricate the
black gasket on the side of the ring. Replace
the ring with the top gasket facing up.
Check the air filter (A) for discoloration
monthly.
If replacing filter, turn off the air source and
bleed the high-pressure hose of air. Loosen
clamps holding the filter and slide it off the
hose. Position the new filter by referencing the
arrow on the filter case toward the back of the
Biostar.