Operation
There are three basic functions to operate the MiniSTAR S including:
Model preparation
•
•
Heating cycle
Pressure molding process
•
Model Preparation
Prepare models for the thermal-forming process by removing bubbles or filling voids whether using the platform or
pellets. The platform is used when forming soft or thin plastics. Pellets are used when forming thick, less flexible
materials. The pellets prevent the plastic from forming to unwanted areas of the dental cast, reducing material thinning.
Platform Use
The platform is used to expose the entire dental mold to the formation of the thermal plastic. It eliminates having to
use pellets. Pellets are difficult to remove from softer material (i.e., soft mouthguard material) and can create a higher
occurrence of air leaks during the pressure-forming process when using thinner thermal plastics (i.e., < 1mm thick).
7
B
A
C C
C
Check the air filter clamped to air line (white arrow) monthly
for discoloration and/or moisture.
If replacing the filter, turn off the air source and bleed the
high-pressure hose of air. Loosen clamps holding the filter
and slide it off the hose. Position the new filter by referencing
the arrow on the filter case toward the back of the MiniSTAR S.
The heating surface (A) should be parallel to the heating
element (B). Adjust the position of the chamber by turning
the screw (C) located under the chamber support frame with
a small screwdriver.
Clockwise adjustment of the screw lowers the chamber from
the heater, counterclockwise raises the chamber closer to the
heater.
Check the position of the chamber when heating materials.