AffloVest RP8300 Petunjuk Penggunaan dan Panduan Pemilik - Halaman 9

Jelajahi secara online atau unduh pdf Petunjuk Penggunaan dan Panduan Pemilik untuk Peralatan Medis AffloVest RP8300. AffloVest RP8300 17 halaman.

AffloVest RP8300 Petunjuk Penggunaan dan Panduan Pemilik
B
F
UTTONS AND
UNCTIONS
I
NTENSITY
Press the P3 button to start the PERCUSSION program.
All motors operate in a pulsating/tapping mode.
Percussion of the upper body with a regular rhythm.
Press the P2 button to start the VIBRATION program.
All motors operate uniformly at the set intensity level.
Deep oscillation of the thoracic cavity.
Press the P1 button to start the DRAINAGE program.
The individual zones are activated in succession.
Traveling oscillation up and down the thoracic cavity.
After starting the programs, the motors run through a "soft start"
function (approx. 10 seconds) up to the set level of intensity.
Programs P1, P2, and P3 always start with SOFT intensity.
Note: The intensity automatically switches to SOFT when changing the programs P1, P2, and P3.
You can then freely select your chosen intensity levels.
If the program is not changed, the vest will switch off automatically after approx. 8 minutes. After each program
change, the new program continues to run for approx. 8 minutes..
You can stop the application at any time by pressing the STOP button.
L
EVELS