3) Operation and Installation

3.1) General

The general sensor connector configuration is given below;
Cable Code/Pin Configuration:
• Red
: V +
• Black
: Ground
• X-Axis: Yellow : Signal(+)
Purple : Signal(-)
• Y-Axis:
Blue : Signal(+)
Green : Signal(-)
• Z-Axis: White : Signal(+)
Orange : Signal(-)
WARNING
Never connect the power supply and/or the power ground to yellow, purple, blue, green,
white, and/or orange cables.
Never connect the power supply to the power ground. Always use a clean power source
and check the voltage range.
4)

Sensor Static Calibration Verification

Using gravity, voltage values are measured in the + and - directions, providing a value of 1 g.
The measurement should be made as follows;
When the acceleration value of 5000 series sensors is entered into the data
acquisition system, the sensor shows +1 g with the effect of gravity, which is
in the direction of the axis to be calibrated.
When the sensor is positioned in the opposite direction to the axis to be
calibrated, the arrow shows -1g as shown below under the effect of gravity.
Using gravity, the voltage values that provide 1 g in the + and - directions
are measured and compared with the catalog value. The calibration value
should be close to the catalog value with a 10% tolerance. Sensor catalog
sensitivity values are given in Table 1.
Power supply voltage +10 V to +40 VDC
Power GND
Positive, analog output voltage signal
Negative, analog output voltage signal
Positive, analog output voltage signal
Negative, analog output voltage signal
Positive, analog output voltage signal
Negative, analog output voltage signal
7