Gako UV-C Panduan Pengguna - Halaman 5

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan Laboratorium Gako UV-C. Gako UV-C 8 halaman. Air sterilizer

6. Functions and Use

6.1 Graduation

No need of extra installation. Electrical connections are fully insulated.

6.2 Operation

• Make sure that all windows and doors are closed during disinfection;
• Plug in the power cord on energy supply;
• Press the Power switch;
• Make sure there are no obstruction on the airflows (sides and front);
• The device is working.

7. Attention

• Please read the instruction manual carefully before using the device;
• The usage must be carried out in accordance with the rules stipulated in the manual;
• It is forbidden to use the device in inflammable and/or explosive places;
• Doors and windows must be closed during disinfection;
• Power must be turned off during installation and maintenance;
• It is strictly forbidden to cover or block the air inlets and outlet of the device;
• Do not insert anything into the device, as it may cause fire or electric shock;
• Do not spill liquids inside the device;
• The power should be turned off before cleaning the device;
• During maintenance or cleaning, turn off the power to avoid UV rays from puncturing the skin and eyes;
• The power supply used by this unit must be a single-phase three-core power supply with good grounding;
• To reduce the risk of electric shock, do not disassemble the unit;
• Please unplug the device from energy supply and contact your local supplier when the following occurs:
A
When the power cord or plug is damaged or worn;
B
If liquid is spilled into the unit;
C
If the sterilizer is being exposed to rain or water;
D
If the sterilizer is broken or the frame is damaged;
E
If the sterilizer changes significantly in performance.
• The service life of the whole device is five years. If used for more than five years, the aging of components and
power cords may cause personal safety accidents. A complete Service must be done after this period;
• Ultraviolet germicidal lamps should be used to detect the irradiance according to the time of use, and the irradi-
ance of the radiation is <70µW/cm2 (power ≥30W). The UV germicidal lamp should be replaced regularly (lifetime
more than 5000 hours).
5